User Guide

Instruction Manual CyberScan pH 11 / 110
14
7. Rinse the probe with de-ionised water or a rinse
solution, and place it in the next pH buffer.
8. Follow steps 5 and 6 for additional calibration
points. See Figure 12.
9. When all the calibration points set in the Unit
Configuration Setup (see P1.3 Select Number
of pH Calibration Points on page 36) are
completed, the meter returns to Measurement
mode automatically. However, if you wish to
terminate the calibration without completing the
number of points as set in the Unit
Configuration Setup menu, press CAL/MEAS to
return to pH measurement mode.
NOTES:
1. To exit from pH calibration mode without confirming
calibration, DO NOT press ENTER in step 6. Press
CAL/MEAS instead.
2. To limit the number of pH buffer values available during
calibration, see P1.3 Select Number of pH Calibration Points
on page 36.
CAL
pH
pH
READY
4.15
4.01
CAL
pH
pH
READY
4.01
4.01
ENTER
HOLD
Figure 12: Proceed to next
calibration
p
oint