User Guide

Instruction Manual CyberScan pH 11 / 110
35
9.2.2 P1.2: Select Calibration Buffer Options
The types of pH calibration buffer options available are:
Meter Meter Display pH Calibration Buffer Options
pH 11 USA
nSt
USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45.
NIST buffers --- pH 1.68, 4.01, 6.86, 9.18, and 12.45.
pH 110 USA
nSt
dn
PUr
USA buffers --- pH 1.68, 4.01, 7.00, 10.01, and 12.45.
NIST buffers --- pH 1.68, 4.01, 6.86, 9.18, and 12.45.
DIN buffers --- pH 1.09, 3.06, 4.65, 6.79, 9.23, and 12.75.
PWB ( Low Ionic): pH 4.10 and 6.97
From measurement mode
Press SETUP key to enter Set Up mode.
1. Press the MI/ or MR/ keys to
scroll through subgroups until you
view parameter P1.0.
2. Press the HOLD/ENTER key two
times to select parameter 1.2.
3. Press the MI/ or MR/ keys to
select the buffer set you require:
4. Press the HOLD/ENTER key to
confirm selection and to return to the
subgroup menu. Press CAL/MEAS
key to return to measurement mode.
See Figure 31.
This screen is
available only
in the pH 110
meter
This screen is
available only
in the pH 110
meter
SETUP
SETUP
usa
p 1.2
p 1.2
SETUP
p 1.2
SETUP
p 1.2
MR
MI
MR
MI
MR
MI
Figure 31: P1.2 - Select pH calibration
buffer standards