User Guide

Instruction Manual CyberScan pH 11 / 110
41
9.5 P4.0 : Auto-Off
In Program P4.0 you can activate the AUTO-OFF option to automatically power off
the instrument 20 minutes after the last key selection. This feature is useful for
conserving battery power. Under default conditions, the instrument has this option
activated.
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/ or MR/ keys to scroll through
subgroups until you view parameter P4.0.
3. Press the HOLD/ENTER key.
4. Press MI/ or MR/ keys to select options
‘YES” - Meter will automatically switch off
20 minutes after the last key operation.
‘NO’ - Auto Off mode is switched off.
5. Press the HOLD/ENTER key to confirm and
return to P4.0.
6. At any point, you can press the CAL/MEAS
key to return to measurement mode.
Figure 37: P4.0 - Auto shut off
feature
ENTER
HOLD
SETUP
P 4.0
MR
MI
SETUP
SETUP
yes