9.6 P5.0 : Reset to Factory Default
This mode lets you reset all parameters to factory
default settings. There are two levels of reset:
Calibration Reset: Reset only the calibration
values. This clears all previous calibrated
values and reset to factory default. See page 68
and 69 for factory default settings.
User Reset: Clears all data including
calibration, memory and other customised
setup functions and reset it to factory default.
See page 68 and 69 for factory default settings.
9.6.1 Calibration Reset
From measurement mode:
1. Press SETUP key to enter Set Up mode.
2. Press the MI/ or MR/ keys to scroll
through subgroups until you view parameter
P5.0.
3. Press the HOLD/ENTER key.
4. Press the MI/ or MR/ keys to toggle
between NO and YES.
‘NO’ retains current settings
‘YES’ erase calibration settings.
5. Press the HOLD/ENTER key to confirm
selection.
If ‘NO’ is selected, it will retains current
settings and proceed to next program
P5.2 – User Reset.
If ‘YES’ is selected, all LCD segments will
lit up for one second, erase calibration
values, and the meter switches to the
measurement mode.
6. Otherwise press CAL/MEAS key to return to
measurement mode without resetting to factory default.
See Figure 38.
yes
SETUP
SETUP
SETUP
SETUP
P 5.0
P 5.1
CAL
CAL
CAL
ENTER
HOLD
ENTER
HOLD
MI
MR
Select either "YES" or "nO",
press ENTER key to confirm.
Figure 38: P5.1 - Calibration
reset