User Guide

Instruction Manual CyberScan pH 11 / 110
70
17 SPECIFICATIONS
MODEL NO. pH 11 pH 110
pH Range -2.00 to 16.00 pH -2.00 to 16.00 pH
Resolution 0.01 pH 0.01 pH
Accuracy ± 0.01 pH ± 0.01 pH
Slope Range 80% to 120% 80% to 120%
Millivolt Range -1999 to +1999 mV -1999 to +1999 mV
Resolution 0.1 mV (between ±199.9 mV)
1 mV (beyond ±199.9 mV)
0.1 mV (between ±199.9 mV)
1 mV (beyond ±199.9 mV)
Accuracy ± 0.2 mV (between ±199.9 mV)
± 2.0 mV (beyond ±199.9 mV)
± 0.2 mV (between ±199.9 mV)
± 2.0 mV (beyond ±199.9 mV)
Relative mV Range - none - + 1999 to -1999 mV
Resolution - none - 0.1 mV (between ±199.9 mV)
1 mV (beyond ±199.9 mV)
Accuracy - none - 0.2 mV (between ±199.9 mV)
2.0 mV (beyond ±199.9 mV)
Temperature Range -10.0 to 110.0 ºC -10.0 to 110.0 ºC (14 to 230 °F)
Resolution 0.1 ºC 0.1 ºC (0.1 °F from 14.0 °F to 199.9 °F ) &
(1 °F from 200 °F to 230 °F)
Accuracy ± 0.5 ºC ± 0.5 ºC (± 0.9 °F)
Temperature Compensation Automatic / Manual (0 to 100 ºC) Automatic / Manual (0 to 100 ºC)
No. Of Calibration Points 2 to 5 points 2 to 6 points
No. Of pH Buffer Values USA: pH 1.68, 4.01, 7.00, 10.01 & 12.45
NIST: pH 1.68, 4.01, 6.86, 9.18, & 12.45
USA: pH 1.68, 4.01, 7.00, 10.01 & 12.45
NIST: pH 1.68, 4.01, 6.86, 9.18, & 12.45
DIN: pH 1.09, 3.06, 4.65, 6.79, 9.23 & 12.75
PWB: pH 4.10 and 6.97
pH Buffer Options USA & NIST USA, NIST, DIN & PWB
Relative mV functions No Yes
Relative mV Offset Display No Yes
pH Slope Display Yes Yes
pH Offset Display Yes Yes
Memory 50 data sets 100 data sets
Outputs None RS232C
Auto Off Selectable Selectable
Auto Off Time 20 minutes after last key operation 20 minutes after last key operation
Averaging/Stability (READY) Selectable Selectable
Display Custom Dual LCD Custom Dual LCD
Hold Function Yes Yes
Auto Hold function No Selectable
Inputs BNC, 2.5 mm phono jack, power jack BNC, 2.5 mm phono jack, power jack
Output No RS232C
Power Requirements 4 ‘AAA’ Batteries;
AC/DC Adapter 9 V, 200 mA
4 ‘AAA’ Batteries;
AC/DC Adapter 9 V, 200 mA
Battery Life > 700 hrs (Alkaline Batteries) > 700 hrs (Alkaline Batteries)
Environmental Requirements 0 to 50 °C (32 to 122 °F) 0 to 50 °C (32 to 122 °F)
eight: & Dimensions Meter: 18.7 x 9 x 4 cm; 290 g
Boxed: 24 x 23 x 7 cm; 650 g
Meter: 18.7 x 9 x 4 cm; 290 g
Boxed: 24 x 23 x 7 cm; 650 g
Note: In strong electromagnetic environment, a deviation of up to 3 % of reading may be noticed.