2.5  Display & Keypad 
2.5.1  Display Overview 
The Liquid Crystal Display (LCD) of the Alpha DO 500 Transmitter has two 
alpha-numerical displays (Upper and a Lower). 
x  Upper display: Measured dissolved oxygen reading (in % saturation or 
ppm or mg/l) is displayed when the Transmitter is in normal operation 
(measurement) mode. 
x  Lower display: Measured temperature value is displayed when the 
Transmitter is in normal operation (measurement) mode. In Calibration 
mode, calibration points are displayed here. 
The two displays indicate function names, options & settings in Setup mode. 
Refer ‘Appendix 2 – Abbreviations Used in LCD’ for more details. 
The LCD also consists of various mode indicators, status annunciators and 
unit of measurement indicators. 
%
SETUP MEAS
CAL
Upper Display
Lower Display
ERR
ATC
°C 
READY
-
-
HOLD
ppm
mg/l
K=
Mode Indicators 
MEAS 
Measurement mode (Refer Section 3.1) 
SETUP 
Setup mode (Refer Section 5) 
CAL 
Calibration mode (Refer Section 4) 
Status Annunciators 
READY 
Appears when the reading is stable 
HOLD 
Appears in Setup mode and Calibration mode to indicate that the relay function 
is frozen 
ATC 
Appears when Automatic Temperature Compensation (ATC) is enabled. 
Not visible when Manual Temperature Compensation (MTC) is enabled. 
Flashes if the temperature probe is faulty in its ATC mode. 
(Refer Section 5.4) 
ERR 
Appears when an error occurs 
Buffer annunciator. Appears in calibration mode or viewing calibrated point 
Electrode annunciator. Appears when viewing electrode properties or during 
calibration error 