Specifications
Instruction Manual    Alpha DO 500 
15 
NOTES: 
x  If the displayed value is more than r10% of the high-point value, the 
Transmitter indicates a calibration error by showing ‘ERR’ and blinking 
electrode 
 annunciator. If this happens, press  and  keys together 
to return to measurement mode. Restart the calibration from Step1. 
x  To exit from calibration mode without confirming calibration, press  
and  keys together. 
x When calibrating with manual temperature compensation, the 
Transmitter automatically changes from preset ‘process temperature’ to 
‘calibration temperature’. After leaving calibration mode, the transmitter 
switches back to the ‘process temperature’. (Refer section 5.4 for 
temperature settings) 
4.2.2 2-Point Calibration 
The 2-point calibration is carried out at Low-point (zero dissolved oxygen 
solution) and then at high-point using 8.24 mg/l of dissolved oxygen solution. 
Measured value should be
<10 % of high-point 
Calibration error occurs
 if measured value is
 <10% of high-point 
during 1st point calibration 
or
if measured value is not
within +/-50% of high-point
during 2nd point calibration 
2
CAL
HOLD
1
ENT
ENT
ENT
CAL
HOLD
2
CAL
HOLD
2
CAL
HOLD
mg/l
MEAS
mg/l
°C
ATC
READY
4
CAL
HOLD
mg/l
READY
ERR
CAL
HOLD
3
mg/l
READY
5
+
ENT
ENT
11
From measurement mode press CAL key to enter calibration mode as 
described in section 4.1. The LCD shows ‘CAL Do’. Press ENT key to 
begin first calibration point. 
2
Select 2-point calibration: The display shows ‘1Pt CAL’. Press  or  
key to select ‘2Pt CAL’. 
Press ENT key to confirm the selection. 
3
The buffer annunciator   appears in LCD. The lower display shows 
‘LO’, indicating the low-point is being calibrated. Immerse the electrode 
in a low level of dissolved oxygen solution (0.0 mg/l). Immerse the 
temperature probe in the solution if ATC is enabled. The upper display 
shows the measured value of dissolved oxygen of the solution. Allow the 
reading to stabilize. LCD shows ‘READY’ annunciator when the reading 
is stable. 
Press ENT key to confirm the reading. 










