Specifications
Instruction Manual    Alpha DO 500 
30 
5.8  Viewing Calibration Point 
Calibration setting lets you view the high-point at which the calibration has been 
performed in the Transmitter. 
ENT
SETUP
HOLD
1
ENT
SETUP
HOLD
2
%
11
From measurement mode press ENT key to enter setup mode as 
described in section 5.1. The LCD shows the first screen of setup mode 
(COF). Press  or  key to  select viewing calibration settings screen 
(CAL). 
Press ENT key to view calibration settings (CAL). 
2
The buffer annunciator 
 appears in the LCD. The upper display shows 
the value of the high-point at which the calibration has been carried out. 
The corresponding unit of measurement of calibration is displayed. 
NOTE: 
If the Transmitter has not been calibrated yet for the currently 
selected unit of measurement, the LCD shows an empty 
reading (“---“ ) in the upper display. 
  Press ENT key to exit calibration settings screen. The Transmitter 
reverts to SET CAL screen. 
Press  or  key to access other setup screens or press  and  key 
simultaneously (escape) to return to measurement mode. 