Specifications
10 
5.2  pH Calibration  
A  complete  2-point  CAL  is  required  for  this  Transmitter.  If  Calibration  is  aborted 
after 1st calibration point, transmitter reverts to previous Calibration data. 
1.  Rinse probe thoroughly with de-ionized water or a 
rinse solution. Blot excess liquid. 
2.  Dip  probe  into  calibration  buffer.  End of  probe 
must  be  completely  immersed  into  buffer.  Stir 
probe gently to create a homogeneous sample. 
3.  From pH MEAS  mode, press CAL to enter pH 
calibration  mode.  CAL  indicator  will  be  shown. 
Primary display will show measured reading while 
secondary display will indicate the pH  standard 
buffer solution. 
4.  Press ▲ or ▼ keys to select either pH 7.00 (USA) 
or  pH  6.86  (NIST)  standard  buffers.  This  is  the  first  point,  Offset 
calibration. 
5.  Wait  for  measured  pH  value  to  stabilize.  Press 
ENTER to confirm calibration. Transmitter is now 
calibrated to current buffer. 
6.  Rinse  probe  with  de-ionized  water  or  rinse 
solution,  and  place  in  next  pH  buffer  (pH  4.01, 
pH 9.18 or pH 10.01). 
7.  Repeat steps 5 and 6. 
8.  Upon successful Calibration, transmitter displays 
Slope  (primary  display,  in  mV)  and  Offset 
(secondary display, in pH) of electrode. 
9.  Press  ENTER  to  return  to  MEAS  mode. 
Transmitter has been calibrated. Rinse electrode 
and proceed with measurements. 
NOTES: 
To  exit  from  pH  calibration  mode  without  confirming  calibration,  press  ▲  and  ▼ 
keys together. 
If  selected  buffer  value  is  not  within  1.0  pH  from  measured  pH  value:  electrode 
and buffer icon blink and ERR annunciator appears in lower left corner of display. 
CAL
HOLD
pH 
READY
pH 
CAL
HOLD
pH 
READY
pH 
CAL
HOLD
pH 
READY
pH 
CAL
HOLD
pH 
mV










