6  ADVANCED SETUP FUNCTIONS 
The  advanced  setup  mode lets  you  customize  your  transmitter‟s preferences  and 
defaults.  This transmitter  features different sub groups  that organize all  setup 
parameters. 
The sub-groups are: 
6.1  Offset Setting (Unavailable in ORP mode) 
 As a result of the continuous measurement and the application concerned, it may 
not be convenient to remove the electrode for calibration. In such cases, an on-line 
Offset adjustment is offered. 
1.  Press ENT key. Transmitter displays „OFS‟. Press 
ENT  again.  Upper  display  shows  amount  of 
offset, in terms of pH; lower display shows current 
measured pH value. 
2.  Use  ▲  or  ▼  keys  to  introduce  offset  to  pH 
readings.  As  the  lower  measured  reading 
changes,  upper  display  records  the amount  of 
offset. Maximum of ± 2.00 pH offset is allowed. 
3.  Press  both  ▲  or  ▼  keys  together  to  return  to  „ 
MEAS „ mode. 