6.2  Output Range Setting 
Output  of  pH  500  transmitter  is  via  the  2-wire  power  supply 
loop.  Both  the  4  mA  and  20  mA  output  can  be  assigned  to 
specific pH or ORP values, for a more refined output. 
1.  Press ENT key and use ▲ or ▼ keys to scroll till LCD 
displays „r n g ‟. Press ENT again.  Upper display 
shows pH value, „0.00 pH‟ (or – 1000 mV) while lower 
display  shows  „r  4.0‟.  This  is the set  up  page  for the 
pH  or  ORP  output  range  setting  for  the  4mA  output 
current. 
2.  Assign pH or ORP value to 4 mA output current. Use 
▲ or ▼ keys to change pH or ORP value. Press ENT 
to accept value. 
3.  Upper display shows pH value,‟14.00 pH‟ (or 1000 mV) while lower display 
shows  „r  20.0‟.  This  is  the  set  up  page  for  the  pH  or  ORP  output  range 
setting for the 20mA output current. 
4.  Assign  pH  or  ORP value  to  20  mA  output  current.  Use  ▲  or  ▼  keys  to 
change pH or ORP value. Press ENT to accept value. 
5.  Press both ▲ or ▼ keys together to return to „ MEAS „ mode. 