6.3  Temperature Compensation Setting  
(Unavailable in ORP mode) 
pH  values  other  than  pH  7.00  are  affected  by  temperature.  Under  varying 
temperature conditions, use ATC to compensate pH values. If temperature of bath 
is  constant,  and  a  temperature  probe  is  not  available,  Manual  Temperature 
Compensation can be utilized. 
6.3.1  Automatic Temperature Compensation 
For automatic temperature compensation (ATC), connect ATC probe to transmitter, 
as described in Section 2.2.2. 
1.  Press ENT  key to enter SETUP  mode. Press ▲ or ▼ keys to scroll 
through sub-menus till LCD displays „SET 
o
C‟. Press ENT. 
2.  Display shows „ATC On‟ or „ATC Off‟. Use  or  keys to select „ATC 
On‟, and press ENT. 
3.  Dip  ATC  probe  into  a  solution  of  known  temperature  (i.e.  a  temperature 
bath).  Allow  time  for  temperature  probe  to  acclimatize  with  surrounding 
bath  temperature.  If  temperature  readings  are  inaccurate,  calibrate 
temperature probe as described in steps 4 and 5. 
If temperature readings are correct, press ENT key and go to step 6. 
4.  Use  or  keys to adjust correct temperature value (i.e. temp. of temp. 
bath).  Readings  can  be  adjusted  in  increments  of  0.1  C.  Maximum 
adjustments allowed is ± 5 
o
C. 
5.  Once you have adjusted the correct temperature, press ENT key. 
6.  Press  and  keys together, to return to MEAS mode. 
The ATC indicator will light up on the LCD. 
NOTE:  If you are using a temperature probe, the probe must be submersed in the 
liquid you are measuring. 
6.3.2  Manual Temperature Compensation 
1.  Make sure the transmitter is on.  Press ENT key and use ▲ or ▼ keys to 
select „SET 
o
C‟. 