Reactivating the pH electrode 
If  stored  and  cleaned  properly,  your  pH  electrode  should be  ready  for  immediate 
use. However, a dehydrated bulb may cause sluggish response.  To rehydrate the 
bulb, immerse electrode in a pH 4 buffer solution for 10 to 30 minutes. If this fails, 
the electrode requires activation. 
pH electrode activation (for glass body electrodes only) 
WARNING:  Only  qualified  persons  proficient  with  the  safe  handling of  dangerous 
chemicals  should  perform  the  procedure  below.  Provide  proper  containers,  fume 
hoods, ventilation, and waste disposal. Safety goggles and protective clothing must 
be worn while performing this procedure. 
1.  Dip or stir the pH electrode in alcohol for 5 minutes. 
2.  Leave the electrode in tap water for 15 minutes. 
3.  Dip and stir the electrode in concentrate acid (such as HCL or H2SO4) for 
five minutes. 
4.  Repeat Step 2. 
5.  Dip and stir in strong base (NaOH) for five minutes.  
6.  Leave for 15 minutes in tap water. 
7.  Now test with standard calibration buffer solutions to see if electrode yields 
acceptable results. You may repeat step 3 through 6 up to three times for 
better  response.  If response does not improve, then your electrode is no 
longer  functioning.  Replace  with  new  electrode  –  call  your  distributor  for 
information. 