8  TROUBLE SH OOTING GUIDE 
Problem 
Cause 
Solution 
Power on but no 
display 
a). Loose connections 
b). Cables not in correct 
polarity (+ and – position). 
a). Check cables are making good 
contact. 
b). Re-wire loop cables with 
correct polarity.  
Unstable readings 
a). Air bubbles in probe. 
b). Dirty probe. 
c). Probe not deep enough in 
sample. 
d). External noise pickup or 
induction caused by nearby 
electric motor.  
e). Severe Reference junction 
fouling. 
a). Tap probe to remove bubbles.  
b). Clean the probe and re-
calibrate. 
c). Make sure sample entirely 
covers the probe sensors. 
d). Switch to Symmetrical mode 
and connect PMP. 
e). Replace probe. 
Slow response 
a). Dirty / Oily probe. 
a). Clean probe. See “Probe Care 
& Maintenance”, Section 7.  