Specifications
4 
2.2.2  To connect temperature probe: 
For  Automatic Temperature  Compensated (ATC)  pH  readings, a  100  Pt RTD 
temperature  probes  can  be  used.  Industrial  pH  electrodes  are  often  offered  with 
ATC, which will typically be bare wires.  
3-wire probe 
1.  Insert Pt 100 sense wire to pin 5 of connector. Tighten screw. 
2.  Insert Pt 100 input wire to pin 6 of connector. Tighten screw. 
3.  Insert Pt 100 return wire to pin 7 of connector. Tighten screw. 
2-wire probe: 
1.  Take a small length of wire and short pins 5 and 6 of connector. Insert Pt 
100 sense wire into pin 6 of connector and tighten screws. 
2.  Insert Pt 100 return wire to pin 7 of connector. Tighten screw. 
IMPORTANT:  Temperature  probe  should  always  be  together  with  pH  probe, 
even during calibration. Make sure no bare wires are exposed and all screws 
are tightened for proper contact. 
2.2.3  To connect Potential Matching Pin (Liquid electrical interferences): 
Connect  the  additional  wire  from  the  electrode  to  pin  8.  Transmitter  must  be 
configured  to  „Symmetrical  mode  of  operation‟  -  Please  refer  to  „Advanced  Setup 
Functions‟,  Section  6.6.2  –  Selecting  Asymmetrical  or  Symmetrical  Mode  of 
Operation. 