Instruction manual
- 17 - 
NOTES: 
•  To exit calibration mode at any time during calibration, press ▲ 
and ▼ keys simultaneously (escape). The pH Monitor returns 
to the measurement mode and the old calibration values remain 
active 
•  The calibration is always carried out in the units of 
measurement (pH or mV), selected in setup mode 
4.3 pH Calibration 
The pH Monitor is capable of calibration of up to 2 points using USA or NIST pH 
buffer standards. All new calibration values will automatically override existing 
calibration data. pH buffer values are referenced to 25 
o
C. Make sure the 
measurement mode is set to pH. (Refer Section 5.5 for switching measurement 
modes). All pH buffer values have window of up to ± 1 pH tolerance during 
calibration. Calibration error can be occurred if the measured pH value exceeds this 
window. If you wish to abort the pH calibration, press ▲ and ▼ keys simultaneously 
and the pH Monitor reverts to pH measurement mode. 
4.3.1  For USA Standard Buffer 
Make sure that the pH Monitor is set to accept USA standard buffer in the Setup 
mode. (Refer Section 5.4 for Buffer Selection Settings) The factory default is USA 
standard. It is recommended that you perform 2-point calibration at room 
temperature (25 
o
C), starting with the first buffer at pH 7.00 followed by any other 
buffer value. (pH 4.01, 1.68, 10.01 or 12.45) 
1-Point Calibration: 
CAL
HOLD
ENT
CAL
pH
READY
HOLD
pH
CAL
CAL
mV
pH
HOLD
MEAS
PH
°C
ATC
READY
ENT
1
2
3
4
11
From pH measurement mode press CAL key to enter calibration mode as 
described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to 
begin calibration. 
2
Place the electrode in pH 7.00 buffer. Immerse the temperature probe in 
the buffer solution if ATC mode is enabled. Immerse the potential matching 
pin in the buffer if symmetrical mode is enabled. The buffer annunciator 
appears in LCD. Lower display shows pH 7.00 (USA standard buffer). 