Instruction manual
- 20 - 
1-Point Calibration: 
CAL
HOLD
ENT
CAL
pH
READY
HOLD
pH
CAL
CAL
mV
pH
HOLD
MEAS
PH
°C
ATC
READY
ENT
1
2
3
4
11
From pH measurement mode press CAL key to enter calibration mode as 
described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to 
begin calibration. 
2
Place the electrode in pH 6.86 buffer. Immerse the temperature probe in the 
buffer solution if ATC mode is enabled. Immerse the potential matching pin 
in the buffer if symmetrical mode is enabled. The buffer annunciator 
appears in LCD. Lower display shows pH 6.86 (NIST standard buffer). 
Upper display shows the current uncalibrated pH reading. Allow the reading 
to stabilize. LCD shows ‘READY’ annunciator when the reading is stable. 
Press CAL key to confirm the reading. 
3
The calibration is completed. The pH Monitor re-calculates electrode 
properties based on the calibration. The new slope (in mV) is shown in the 
upper display and pH reading at 0mV (pH7.00 ± offset) is shown in the 
lower display. 
Press ENT key to exit from the calibration. 
4
The pH Monitor reverts to pH measurement mode. 