Instruction manual
- 21 - 
11
From pH measurement mode press CAL key to enter calibration mode as 
described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to 
begin first calibration point. 
2
Place the electrode in pH 6.86 buffer. Immerse the temperature probe in the 
buffer solution if ATC mode is enabled. Immerse the potential matching pin 
in the buffer if symmetrical mode is enabled. The buffer annunciator 
appears in LCD. Lower display shows pH 6.86 (NIST standard buffer). 
Upper display shows the current uncalibrated pH reading. Allow the reading 
to stabilize. LCD shows ‘READY’ annunciator when the reading is stable. 
Press ENT key to confirm the reading. 
3
The pH Monitor moves to the second calibration point. The lower display 
shows next standard buffer value (pH 4.01). Use ▲ and ▼ keys to select 
your second buffer from one of the preset values: pH 4.01 or 1.68 or 9.18 or 
12.45. Remove the electrode from the first buffer, wash and then immerse it 
into the second buffer. Allow the reading to stabilize. LCD shows ‘READY’ 
annunciator when the reading is stable. 
Press ENT key (or CAL key) to confirm the reading. 
4
The calibration is completed. The pH Monitor re-calculates electrode 
properties based on the calibration. The new slope (in mV) is shown in the 
upper display and pH reading at 0mV (pH7.00 ± offset) is shown in the 
lower display. 
Press ENT key to exit from the calibration. 
55
The pH Monitor reverts to pH measurement mode. 
NOTES: 
•  To exit from pH calibration mode without confirming calibration, press 
▲ and ▼ keys together 
•  When confirming the buffer measurement, if the measured pH value is 
not within ±1.00pH from selected buffer value, the electrode 
annunciator 
 blinks and ERR indicator appears in the display. This 
error can also occur if non-standard buffers are used or the electrode 
has worn out. If this happens, press both ▲and ▼ keys together to 
restart the calibration beginning from Step1 
When calibrating with manual temperature compensation, the meter automatically 
changes from the preset process temperature to the calibration temperature. After 
leaving the calibration mode, the pH Monitor reverts back to process temperature. 
4.4 ORP Calibration 
The pH Monitor allows a single point ORP calibration. ORP calibration process is 
available only when the measurement mode is set to ORP. (Refer Section 5.5 for 
switching measurement modes). 










