Instruction manual
- 22 - 
CAL
HOLD
ENT
CAL
mV
READY
HOLD
CAL
mV
HOLD
MEAS
°C
ATC
READY
ENT
1
2
3
4
READY
R.mV
11
From ORP measurement mode press CAL key to enter calibration mode 
as described in section 4.2. The LCD shows ‘CAL OrP’. Press ENT key to 
begin calibration. 
2
Place the electrode in a standard ORP solution of known OPR value. 
Immerse the temperature probe in the buffer solution if ATC mode is 
enabled. Immerse the potential matching pin in the buffer if symmetrical 
mode is enabled. The buffer annunciator 
 & electrode annunciator   
appear in LCD. The upper display shows the current mV output of the 
electrode with offset adjustment (if any). The lower display shows the 
default (uncalibrated) mV output of the electrode. Allow the reading to 
stabilize. LCD shows ‘READY’ annunciator when the reading is stable. 
Press ▲ or ▼ key to set the upper display value to the ORP value of the 
solution. Allowable range: -150mV to 150mV. 
3
Press ENT key to confirm the value. 
4
The pH Monitor reverts to ORP measurement mode. 
4.5 Temperature Calibration 
Calibrate temperature probe only if temperature value displayed on the pH Monitor 
is different from that of a calibrated thermometer. Refer to section 5.3 for further 
information temperature settings. 