5.2 Electrode Offset Settings 
NOTE: 
‘Electrode Offset Setting’ is not available when the pH Monitor is 
configured for ORP measurement mode. (Refer Section 5.5 for 
switching measurement modes) 
In applications where continuous pH measurement is required, it may not be 
convenient to remove the electrode for calibration. In such cases, an on-line offset 
adjustment is recommended. The pH Monitor allows you set an offset of up to ± 
2.00pH to compensate for errors in the pH electrode. 
The pH Monitor adds or subtracts the offset value from the measured pH value and 
displays the corrected value. However, if you need to offset the value beyond the 
average offset you would expect in your application type, consider a full calibration 
or even electrode replacement. 
To Determine the Offset Value: 
1.  Take a sample from the liquid of pH measurement. Record the pH reading 
of the pH Monitor at the time the sample was taken. 
2.  Measure the pH value of your sample using another calibrated pH meter 
having its own electrode (pH tester, hand-held meter or bench meter) 
3.  The difference between these two readings is the offset 
SETUP
ENT
ENT
1
pH
READY
pH
SETUP
pH
READY
pH
SETUP
2
3
11
From pH measurement mode press ENT key to enter setup mode as 
described in section 5.1. The LCD shows the first screen of setup mode 
(OFS). Press ENT key to access electrode offset setting (OFS). 
2
The upper display shows the last configured offset value (if any), otherwise 
zero. The lower display shows currently measured pH reading (including 
last configured offset value). Allow the reading to stabilize. LCD shows 
‘READY’ annunciator when the reading is stable. 
Press ▲ or ▼ key to set the upper display to the newly calculated offset 
value. 
3
As the upper display value (offset) changes, the pH Monitor adjust the lower 
reading automatically to suit the new offset value. Up to ± 2.00 pH offset is 