Instruction manual
- 25 - 
allowed. 
Press ENT key to confirm the value. The pH Monitor reverts to OFS 
screen. 
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key 
simultaneously (escape) to return to measurement mode. 
5.3 Temperature Settings  
Automatic Temperature Compensation (ATC): 
The pH values other than pH 7.00 are affected by temperature. Use ATC feature of 
the pH Monitor to compensate for pH changes when the temperature of the sample 
or process liquid fluctuates. Connect a separate temperature probe (Pt100/pt1000) 
to the pH Monitor if an integrated pH/Temperature probe is not available. 
Manual Temperature Compensation (MTC): 
Set the pH Monitor to MTC (disable ATC) when the temperature of sample or 
process liquid is constant and a temperature probe is not available. If you disable 
ATC, the pH Monitor allows you to set your process temperature (P.ºC) and 
calibration temperature (C.ºC). This allows calibration at a different temperature 
other than the process temperature. Example: Setting calibration temperature of 
25°C lets you calibrate using standard solutions at 25°C, even if your process 
temperature is different from 25°C. 
ENT
1
SETUP
2
SETUP
°C
2
ENT
ENT
3
SETUP
°C
SETUP
°C
3
ENT
SETUP
°C
A
TC
SETUP
°C
ATC
5
4
SETUP
ENT
6
SETUP
7
ENT
ENT
ENT
°F
SETUP
11
From pH or ORP measurement mode press ENT key to enter 
setup mode as described in section 5.1. The LCD shows the first 
screen of setup mode (OFS). Press ▲ or ▼ key to select 
Temperature settings screen (SET ºCF). 
Press ENT key to access temperature settings (SET ºCF). 