Instruction manual
- 27 - 
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key 
simultaneously (escape) to return to measurement mode. 
To exit from any intermediate steps, press  ▲ and ▼ keys simultaneously 
(escape). The pH Monitor returns to the first screen of temperature settings SET 
ºCF. 
5.4  Buffer Selection Settings 
NOTE: 
The buffer selection setting is only available if the monitor is 
configured for pH measurement mode. 
This allows you to set the pH Monitor to accept either USA or NIST pH buffer 
standard group during pH calibration. 
Buffer Group  pH Buffer Points 
USA  pH 4.01, 1.68, 7.00, 10.01 & 12.45 
NIST  pH 4.01, 1.68, 6.86, 9.18 & 12.45 
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
ENT ENT
ENT
1
2
2
11
From pH measurement mode press ENT key to enter setup mode 
as described in section 5.1. The LCD shows the first screen of 
setup mode (OFS). Press ▲ or ▼ key to select buffer settings 
screen (bUFF). 
Press ENT key to access buffer settings (bUFF). 
2
Selecting buffer group: The lower display shows the last 
configured buffer group. Press ▲ or ▼ key to select the required 
buffer group (‘USA’ or ‘nST’). 