8 TROUBLESHOOTING 
Problem  Cause  Solution 
Power on, but no display  a)  Loose connections 
b)  Incorrect output voltage of 
the power adaptor 
a)  Ensure cables make good 
contact 
b)  Use an power adaptor with 
specified output voltage 
Unstable pH reading  a)  Dirty electrode 
b) Electrical noise 
interference 
a)  Clean electrode and 
recalibrate 
b)  Switch to Symmetric mode 
Oscillating temperature 
readings 
a) Electrical noise 
interference 
a)  Ensure shield wire is 
properly connected to pin 7 
Slow response  a) Dirty / Oily electrode  a) Clean electrode 
Blinking ATC  a)  No temperature probe 
connection during ATC 
mode 
a)  Ensure temperature sensing 
cable makes good contact 
Blinking electrode 
annunciator 
a)  Error in calibration  a)  Ensure calibration standard 
solution is not contaminated. 
Ensure electrode is clean. 
Or (pH)  a)  pH electrode is not 
connected 
a)  Ensure electrode makes 
good contact with 
Transmitter 
Or (Temperature)  a)  Temperature probe is not 
connected when ATC 
enabled 
a)  Ensure electrode makes 
good contact with 
Transmitter 