Specifications

FM/FMR ZSME300000103
17
Transmission Format
RS1-00 0 ~ 3
S
T
,
G
S
,
+
1
2
3
4
.
5
6
g
CR
LF
Header 1 Header 2 Weight Data (8 digits) Unit Terminators
Header 1
ST: Stable Weight / US: Unstable Weight / OL: Weight Overload
Header 2
GS: Gross Weight / NT: Net Weight / TR: Tare
Weight Data (8 digits)
The first digit of weight data represents “+ / -“indication for weight value. The other 7
digits, including decimal point, represent the weight value. If the weight is over load
(Header 1: OL), the screen turns into “blank” except “+ / -“indication and decimal point.
Unit
kg, lb, t or blank
Terminators
CR and LF are data termination code.
RS1-00 = 4
T
N
,
1
2
3
CR
LF
RS1-00 = 5
RS1-00 = 6
T
W
,
+
1
2
3
4
.
5
6
k
g
CR
LF
D
A
T
E
:
2
0
X
X
/
X
X
/
X
X
CR
LF
T
I
M
E
:
X
X
:
X
X
:
X
X
CR
LF
G
R
O
S
S
:
+
1
2
3
4
.
5
6
k
g
CR
LF
N
E
T
:
+
1
2
3
4
.
5
6
k
g
CR
LF
T
A
R
E
:
+
1
2
3
4
.
5
6
k
g
CR
LF
T
N
:
X
X
X
CR
LF
T
W
:
+
1
2
3
4
.
5
6
k
g
CR
LF