Specifications

FM/FMR ZSME300000103
23
6-2 Weight Calibration
Turn on and warm up the machine for 15 to 30 minutes before calibration.
Adjust the calibration switch to ON, and the screen will show .
Press or to select .
Procedure
Zero Calibration
a) Ensure nothing on the platter; after being stable, press , and the screen will display
……. ”. The zero calibration will be complete 5 seconds later.
b) To abort zero calibration, just press .
Weight Calibration
a) Place an object, whose weight is known, on the platter, and input the weight value from
front panel. After being stable, press and the screen will show ……. ”. The
weight calibration will be complete 5 seconds later.
b) To abort weight calibration, just press .
Zero calibration completed
Confirmed
Set calibration
switch to OFF
Place the
weight
mass and
set the
weight
ENTER
ENTER
ENTER
ENTER
TARE
ZERO
Zero
Calibration
Weight
Calibration
ENTER
ENTER
ESC
ESC
ESC
ESC