Specifications
FM/FMR ZSME300000103
26
Internal Function Parameter Description
Parameter
Code
Function
Setting Value
Default
Setting
Parameter
Description
CFN-00
Tare or Zero
Function under
Unstable Status
0
ON
1
1
OFF
CFN-01
Re-zeroing after
Turning on
0
OFF
1
1
ON
CFN-02
Re-zero Range
0% ~ 30%
0%: Full range re-zero
1% ~ 30%:
Capacity setting value%
2
CFN-03
Filter Strength
0 ~ 5
Strength increases by number
2
CFN-04
Animal scale
Sampling Rate
0
No limit
0
1
20 times/s
2
10 times/s
3
5 times/s
CFN-05
Animal Scale
Mode
0
OFF
0
1
Mode 1: No weight shows under
unstable status
2
Mode 2: Weight shows no matter
under stable or unstable status
CFN-06
Animal Scale
Stable Range
0 ~ 100
Mode 2: Stable Range Setting
30
CFN-07
Animal Scale
Sampling
Frequency
0
8 times
2
1
16 times
2
32 times
3
64 times
4
128 times
CFN-08
Dual Range
Resolution
Setting
0
Multi - interval
0
1
Multi - range
CFN-09
A / D
sampling Rate
0 ~ 9
Strength increases by number
4
CFN-10
Dual Range
Resolution
Middle Point
Setting
0 ~ 999 999
Set point of dual-range
resolution
5 000
CFN-11
G Value
Calibration
9.78032
|
9.83218
Adjust G value
9.79585