2-4 WEIGHT CALIBRATION         
In the weighing mode, press the NET/GROSS  and ON/ZERO  keys at the same time 
to enter the Advanced Function setting mode.  The LCD shows           and 
use the NET/GROSS  or UNITS  key to select         to enter the weight 
calibration mode. 
4 In approved models,            set as 1 or 3, then         is disabled. 
4  Weight calibration conditions: 
The calibration weight value placed on the weight pan must be over e100, and the 
standard deviation of the weight must be within 10%. 
The LCD displays the calibration 
weight value (the maximum capacity 
of 1
st
 unit) and the unit.  The digit on 
the right side keeps flashing. 
The scale is calculating the internal 
value of the zero point.  After 
stabilization, the LCD shows the 
calibration weight value. 
Press 
 TARE 
 key 
When the digit on the right 
flashes, ensure that the 
weigh pan 
is empty.  Then 
Press the 
 TARE 
key. 
After stabilization
Place the displayed weight 
on the weigh pan. Press the
TARE 
 key. The scale 
calculates the full capacity 
internal value and after 
stabilization, the buzzer 
beeps 3 times. Remove the 
weight from 
the weigh pan 
and the procedure 
is completed. 
       .     
   .     
ACTIONS 
DISPLAY 
NOTE 
ON/ZERO 
 key 
⇒
 Upward key (from 0 to 9) 
UNITS 
 key ⇒  Downward key (from 9 to 0) 
TARE 
 key ⇒ Move the cursor to right 
NET/GROSS 
 key ⇒ Move the cursor to left. 