Owner's Manual
7
Step 1
Slide across forehead.
Place probe ush on
center of forehead and
depress button. Keeping
button depressed slowly
slide probe mid-line across
forehead to the hair line.
Measure only the up-side
on a patient in a lateral
position. The down-
side will be insulated
preventing the heat from
dissipating, resulting in
falsely high readings.
Step 2
Slide behind ear.
Keeping button depressed, lift
probe from forehead, touch
behind ear halfway down the
mastoid process and slide
down to the soft depression
behind the earlobe.
Think of a sweatband.
Measure straight across the
forehead and not down the
side of the face. At mid-line,
the temporal artery is about 2
mm below the surface, but can
go deeply below the surface
on the side of the face.
Step 3
Release button, remove
from head and read.
Minimum measuring time:
2 seconds. Minimum time between successive measurements: 30 seconds
Measure exposed skin.
Brush the hair and bangs
aside if covering the area to
be measured.
3-Step Adult Temperature Measurement
How to improve the accuracy of your measurements on adults.
Exergen TAT-5000S