User guide

Remote Control
122 FLS-2600B
Programming Commands
SOURce:SWEep:STEPped:PROGram
Description This function is used to set the parameters for a step-by-step
sweep program, during which the application sweeps between
two wavelengths, stopping at preset intervals.
Syntax SOUR:SWE:STEP:PROG<wsp>[<start>],[<stop>],
[<pause>], [<step>],[<cycles>],[<rev>],[<trig>]
Parameters The <start> and <stop> parameters represent the start and
stop wavelengths for the sweep. The current spectral
measurement units apply (nm or THz). The format must be
9999.999 nm or 999.9999 THz.
The <pause> parameter is the pause length between the steps.
The format must be 59:999.
The <step> parameter is the size of the step between the
measurements. The format will be 9999.999 nm or
999.9999 THz.
The <cycles> parameter is the number of cycles to be
performed during the sweep. The format will be 1-99 or 0
(continuous).
The <rev> parameter indicates whether the reverse option is
on or off. The format must be 1/0.
The <trig> parameter indicates whether the trigger option is on
or off. The format must be 1/0.