User guide

Remote Control
Tunable Laser Source 123
Programming Commands
Note You cannot use this command during a sweep.
You do not have to enter all values in the command, but you
must enter the values in the same order as in the Syntax section.
If you want to enter only some of the values, use a comma ( , )
to separate the values, including the ones you did not enter.
For each value including units, such as wavelengths or speed,
you can specify a unit to use in the same way as the specific
queries in the previous pages. If you do not specify the units, the
system will use those that are currently set.
To specify the units, add <wsp><unit> at the end of the value
you want to modify. The <unit> parameter will be a wavelength
unit (nm or THz) or a speed unit (nm/s or THz/s).
Example A complete command:
SOUR:SWE:STEP:PROG 1570.00,1605.50,00:050,5.500,10,1,0
A command changing only the number of cycles and trigger
options:
SOUR:SWE:STEP:PROG ,,,,5,,1
A command setting the start and end wavelengths in THz, as
well as the pause time:
SOUR:SWE:CONT:PROG 196.7 THZ,186.8 THZ,00:100
SOURce:SWEep:STEPped:PROGram