User guide

Remote Control
124 FLS-2600B
Programming Commands
SOURce:SWEep:STEPped:PROGram?
Description This function is used to set and/or view the parameters for a
step-by-step sweep program, during which the application
sweeps between two wavelengths, stopping at preset intervals.
This function also returns the validated parameters.
If you only wish to view the parameters without programming
them, simply enter the query without any data following it.
Syntax SOUR:SWE:STEP:PROG?<wsp>[<start>],[<stop>],
[<pause>],[<step>],[<cycles>],[<rev>],[<trig>]
SOUR:SWE:STEP:PROG?
Parameters The <start> and <stop> parameters represent the start and
stop wavelengths for the sweep. The format must be
9999.999 nm or 999.9999 THz.
The <pause> parameter is the pause length between the steps.
The format must be 59:999.
The <step> parameter is the size of the step between the
measurements. The format will be 9999.999 nm or
999.9999 THz.
The <cycles> parameter is the number of cycles to be
performed during the sweep. The format will be 1-99 or 0
(continuous).
The <rev> parameter indicates whether the reverse option is
on or off. The format must be 1/0.
The <trig> parameter indicates whether the trigger option is on
or off. The format must be 1/0.