User guide

Remote Control
138 FLS-2600B
FLS-2600B Error and Warning Messages
61 A problem occurred with the
thermo-electric cooler.
Contact EXFO.
62 The Flash memory was not
detected by the tunable laser
source.
Contact EXFO.
63 The Flash reading was not
performed in the set delays.
Contact EXFO.
64 The EEPROM was not
detected.
Contact EXFO.
65 A checksum error was
detected by the Tunable
Laser Source.
Contact EXFO.
66 A problem occurred while
the motor was moving.
Contact EXFO.
67 The Tunable Laser Source
index was lost.
Restart the Tunable Laser
Source.
68 The Tunable Laser Source is
outside its temperature
settings.
Ensure that the ambient
temperature is within the
Tunable Laser Source’s range.
69 An FPGA problem was
detected.
Contact EXFO.
71 An error occurred during a
sweep.
Retry your sweep.
72 An EEPROM error has been
detected.
Contact EXFO.
100 A command could not be
added to the command pipe.
Avoid turning the selection dial
while the Tunable Laser Source
is performing a sweep.
Error or
warning
Number
Description Recommended Action