User guide

Safety Information
8 FLS-2600B
Electrical Safety Information
Electrical Safety Information
The following safety precautions must be observed while operating and
servicing the unit. Failure to comply with these precautions or with specific
indications elsewhere in this manual violates safety standards of intended
use of the unit. EXFO assumes no liability for the user’s failure to comply
with these requirements.
This unit is intended for indoor use only.
Only fuses with the required rated current and specified type (IEC,
250 V, 2 A, fast blow, 0.197 in x 0.787 in/5 mm x 20 mm) may be used
for replacement. Do not use repaired fuses or short-circuited fuse
holders.
The power cable of the FLS-2600B Tunable Laser Source is the most
effective method to turn off the unit if a problem should occur.
Capacitors inside the unit may be charged, even if the unit has been
disconnected from its electrical supply.
Before turning on the unit, all grounding terminals, extensions cords,
and devices connected to it should also be connected to a protective
ground via a ground socket. Any interruption of the protective
grounding is a potential shock hazard and may cause personal injury.
Whenever the ground protection is impaired, the unit is not to be used,
and must be secured against any accidental or unintended operation.
Unit covers cannot be removed during operation.
Any adjustments, maintenance, and repair of opened units under
voltage should be avoided and carried out only by skilled personnel
aware of the hazards involved. Do not attempt internal service or
adjustment unless a person qualified to perform first aid is present.
Do not replace any components while power cable is connected.