User guide

Maintenance
50 FLS-2600B
Replacing the Fuse
Replacing the Fuse
The FLS-2600B contains two fuses of type IEC, 250 V, 2 A, fast blow
0.197 in x 0.787 in/5 mm x 20 mm. The fuse holder is located at the back of
the FLS-2600B, just beside the power inlet.
To replace the fuses:
1. Unplug the power cord from the FLS-2600B.
2. Pull the fuse holder out of the FLS-2600B.
3. Check and replace the fuses if necessary.
4. Make sure the fuses are placed firmly in the holder prior to its
reinsertion.
5. Firmly push the holder into place.
Use only with a 250V fuse
Employer uniquement avec
un fusible de 250V