User guide

Remote Control
Tunable Laser Source 75
Communication Parameters
Communication Parameters
Note: EOS means “End of String”. EOI means “End or Identify”.
For GPIB Communication
Ter minate Read on EOS Yes
Set EOI with EOS on Writes Yes
Type of compare on EOS 8 bits
EOS byte 0Ah
Sens EOI at end of Writes Yes
GPIB primary address See Setting the FLS-2600B for
Remote Control on page 71
GPIB secondary address None
For RS-232 Communication
EOS bytes 0Ah
Baud rate 1200/2400/4800/9600/
19200 bps
Parity None
Data bits 8 bits
Stop bits 1 bit
Flow control Software (xOn/xOff) or
None
Activation See Setting the FLS-2600B
for Remote Control on
page 71