CALibration:WAVelength
Description This command introduces an offset into the FLS-2600B to
correct a wavelength deviation as measured by a wavelength
meter.
Syntax CAL:WAV<wsp><wavelength>
Parameters The <wavelength> parameter is a value indicating the
wavelength in nm.
Example CAL:WAV 1560.000
Note This function cannot be performed if the application is
performing a sweep.
The difference between the wavelength and the measured
wavelength cannot be greater then ± 0.200 nm or ± 0.0250 THz.
EXFO recommends using 1560 nm as the wavelength
parameter.