Operating Your Chromatic Dispersion Analyzer
Chromatic Dispersion Analyzer 29
Performing Reference Measurements
Reference
type
Recommended reference
frequency
Remarks
Wavelength Before each test session
and every time the
application recommends it.
³ Adjusts the source wavelength precisely.
³ The application will advise you to take a
wavelength reference whenever the
need arises (after selecting a different
source type, after module initialization,
after a certain period of time or when
temperature varies), but never during
acquisitions.
When references are necessary, a
message will appear in the status bar,
recommending a new reference.
Chromatic
dispersion
Once every month and
every time the source or
patchcord is changed.
³ Used to cancel out source, analyzer and
patchcord dispersion.
It will be taken according to the sources
you have selected
(C-source; C-reference).
³ Important for low-chromatic-dispersion
measurement.