User guide

SCPI Command Reference
74 IQS-2600B
Product-Specific Commands—Description
Example(s) SOUR:POW:STATE ON
- Wait for the power on operation to be complete.
SOUR:WAV:CW 1560E-009
- Wait for one hour.
- To recalibrate the complete wavelength range,
EXFO recommends using 1560.000 nm as the
reference.
- In this example, the output wavelength
measurement is taken. Supposing that the
measurement wavelength is 1560.023, then:
CAL:WAV 1560.023E-009
Notes This function cannot be performed while the
module is sweeping.
The difference between the current wavelength
and the measured wavelength cannot be greater
then ± 0.200 nm.
EXFO recommends to calibrate the module at
1560 nm.
:CALibration[1..n]:WAVelength