Function Range Resolution   Accuracy 
Frequency 
5.000Hz 0.001Hz 
(1.5% reading + 5 digits) 
50.00Hz 0.01Hz 
500.0Hz 0.1Hz 
(1.2% reading + 2 digits) 
5.000kHz 0.001kHz 
50.00kHz 0.01kHz 
500.0kHz 0.1kHz 
5.000MHz 0.001MHz 
(1.5% reading + 4 digits) 
10.00MHz 0.01MHz 
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms min @ 20% to 80% duty 
cycle and > 100kHz. 
Duty Cycle  0.1 to 99.9%  0.1% 
(1.2% reading + 2 digits) 
Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz 
Temp 
(type-K) 
-20 to 750C  1C  (3.0% reading + 3 digits) 
(probe accuracy not included) 
-4 to 1382F 1F 
NOTE: Accuracy specifications consist of two elements: 
  (% reading) – This is the accuracy of the measurement circuit. 
  (+ digits) – This is the accuracy of the analog to digital converter. 
Diode Test  Test current of 0.3mA maximum, open circuit voltage 1.5V DC typical 
Continuity Check  Audible signal will sound if the resistance is less than 150 (approx.), 
test current <0.7mA 
Temperature Sensor  Requires type K thermocouple 
Input Impedance >7.5MΩ (VDC & VAC) 
AC Response  True rms 
ACV Bandwidth  50Hz to 1kHz 
Display  4000 count backlit liquid crystal 
Overrange indication  “OL” is displayed 
Auto Power Off  15 minutes (approximately) 
Polarity  Automatic (no indication for positive); Minus (-) sign for negative 
Measurement Rate  2 times per second, nominal 
Low Battery Indication  “ ” is displayed if battery voltage drops below operating voltage 
Battery  One 9 volt (NEDA 1604) battery 
Fuses  mA, µA ranges; 0.5A/250V fast blow 
A range; 20A/250V ceramic fast blow 
Operating Temperature  5ºC to 40ºC (41ºF to 104ºF) 
Storage Temperature -20
o
C to 60
o
C (-4
o
F to 140
o
F) 
Operating Humidity  Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 40ºC(104ºF) 
Storage Humidity  <80% 
Operating Altitude  2000 meters (7000ft) maximum. 
Weight  342g (0.753lb) (includes holster). 
Size  187 x 81 x 50mm (7.36” x 3.2” x 2.0”) (includes holster) 
Safety  For indoor use and in accordance with the requirements for double 
insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage 
Category III 600V and Category II 1000V, Pollution Degree 2. 