Installation Guide
DO600-en-GB_V7.4 12/16
6
AltitudeCompensation
1. WiththeunitON,momentarilypresstheCAL/RECALLbuttontwiceinquicksuccession
(‘SAL’isshowninthelowertemperaturedisplay.
2. PressandholdCAL/RECALLagainfor2secondstoentertheAltitudeMode(‘Ald’willbe
showninthelowertemperaturedisplay).
3. Thefactorydefaultvalueis
sealevel.EachpressoftheMODE/HOLDbuttonincreasesthe
compensationby1000ft.Themaximumvalueis20presses(20,000ftabovesealevel).
4. MomentarilypresstheCAL/RECALLbuttontosavethecompensationsettingandreturnto
thenormalmeasurementmode.
Auto‐PowerOFFFeature
Theautopowerofffeatureautomaticallyshutsthemeteroff10minutesafterthemostrecent
buttonpress.Todisablethisfeature,refertotheDisableAuto‐PowerOFFsection.
DisableAuto‐PowerOFFFeature
WiththeunitON,momentarilypresstheCAL/RECALLbutton,thenquicklypress andholdboth
theMODE/HOLDandON/OFFbuttonsuntil‘off’isdisplayed.TorestoretheAutoPowerOff
Feature(autopowerOFFenable)simplyturnthemeteroffandonagainusingtheON/OFF
button.
LowBatteryIndication
Whenthebatteryvoltagefallsbelowtheoperatingthreshold,‘BAT’willappearonthedisplay.
RefertotheMaintenancesectionforbatteryreplacementinformation.
StoringReadings
1. PresstheMODE/HOLDbuttontostoreareading.Thestoragelocationnumberwillbe
displayedonthelowerdisplay,whilethemaindisplayshowsthestoredreading.Themeter
willentertheHOLDmodeandthe“HOLD”indicatorwillappearontheLCD.
2. PresstheMODE/HOLDbuttonagaintoexittheHOLDmodeandreturntonormal
operation.ThenexttimeMODE/HOLDismomentarilypressedanotherreadingisstored
andsoon.
3. Ifmorethan25readingsarestored,previouslystoredreadings(startingwithreading
number1)areoverwritten.
RecallingStoredReadings
1. MomentarilypresstheCAL/RECALLbuttonandthenwithin4secondsmomentarilypress
theMODE/HOLD.Thelaststoreddatapointlocationwillbedisplayed(1to25).Eachtime
theMODE/HOLDbuttonismomentarilypressedthenextmostrecentlystoreddatapoint
willbedisplayed.
2. Afterthelast
datapointstoredisdisplayed,pressingtheMODE/HOLDbuttonagainreturns
thedisplaytothebeginningofthelist.
3. PressingtheCAL/RECALLbuttonatanytimestopsthedataretrievalprocessandreturnsthe
metertothenormalmeasurementmode.