Product Manual

407860 V2.3 7/09 6
Memory Full Indication
The meter produces a continuous tone when the internal memory has been filled to
capacity (500 readings). The data can be viewed, cleared, or downloaded to a PC.
Viewing Stored Data
1. With the meter OFF, Press and hold the HOLD and REC buttons.
2. Press the POWER button.
3. When the display indicates the existing sample rate, release the HOLD and REC
buttons.
4. The HOLD and the FUNCTION buttons are used to scroll through the stored data.
When either button is pressed, the data reference number will briefly appear and then
the reading for the given memory location will be displayed. The HOLD button
increments the data location number and the FUNCTION button decrements the data
location number.
PC Interface
The meter is equipped with an RS-232 serial data port (located at the top of the meter next
to the probe input jack). This interface was designed to operate with the 407001A Extech
software package which includes two programs and the required meter to pc cable.
1. 407001
A
- Extech Data Acquisition Software: This software allows the user the meter
to a PC and view readings in real-time as they are taken (this software does not
download the contents of the meter’s memory to the PC)
2. DL2005 - Datalogger Download Software: This software allows the user to download
the contents of the meter’s memory to the PC
Download Stored Data to PC (requires DL2005 software)
The SEND button is used to send the contents of the meter’s internal datalogger memory
to the PC via the optional DL2005 software.
1. Connect the RS-232 cable to the PC
2. Install and run the DL2005 datalogging software.
3. With HOLD and REC not selected press and hold the SEND button for > 2 seconds.
“232” (flashing) will appear in the display.
4. Press the SEND button again and the data will be sent to the meter’s RS-232 port
and onto the connected PC.
5. Press the ESC button to return to normal operation.