User Manual

42280 V3.0 10/2007
14
Troubleshooting
Power is ON but there’s no display
Check that the batteries or the AC adaptor is installed and connected properly
Ensure that the START-STOP button is pressed firmly and for at least 100ms
Calibration Failure
Check that the batteries or the AC adaptor is installed and connected properly
Ensure that the calibration bottles are tightly sealed around the meter sensor
Ensure that calibration takes place at room temperature
Flashing Indicators
Display flashes in logging mode when the AC adaptor has become unplugged
DP flashes when temperature is over limit
Degrees F icon flashes when Humidity is over limit
REC flashes when datalogging takes place
ALM light flashes when alarm limit has been exceeded
Error Codes
E02 - Under flow error
E03 - Over flow error
E04 - E02 and E03 combined error
E11 - CAL error (recalibration required)
E32 - IC read/write error. Return unit for repair
E33 - Circuit Error. Return unit for repair