User manual

DA50-EU-V2.0 12/08
User’s Guide
DA50 Non-Contact Current Detector
Safety
WARNING: Risk of Electrocution. Before use, always test the Current
Detector on a known current carrying circuit to verify proper
operation
WARNING: Risk of Electrocution. Keep hands and fingers on the
handle of the probe and away from the probe tip
WARNING: If the equipment is used in manner not specified by
manufacturer, the protection provided by the equipment may be
impaired
CAUTION: Read, understand and follow Safety Rules and Operating
Instructions in this manual before using this product.
Do not attempt to repair this unit. There are no user serviceable parts.
Do not expose the unit to extreme temperatures or high humidity
Do not use the unit if it is wet or damaged
Do not apply more than the rated voltage between the probe tip and
ground.
Do not operate with the case open
This symbol, adjacent to another symbol or terminal, indicates
the user the user must refer to the manual for further
information.
This symbol, adjacent to a terminal, indicates that, under normal
use, hazardous voltages may be present
Double insulation
Specifications
Current Sensitivity 200mA AC at 0.2”
Bandwidth 50/60Hz
Operating Temperature 32 to 122
o
F (0 to 50
o
C)
Storage Temperature -4 to 140
o
F (-20 to 60
o
C)
Altitude Operating below 2000 meters
Relative Humidity 80% up to 31
o
C, decreasing to 50% at 50
o
C
Battery 2 x 1.5V button batteries (LR44/AG13)
Dimensions/Weight 7.7x2.4x1.2”(195x62x30mm)/ 5.2oz (148g)
Safety For indoor use and in accordance with
Overvoltage Category IV 600V, Overvoltage
Category III 1000V, Pollution Degree 2.
Description
1. Current sensor
2. ON and Current indicator lights
3. Sensitivity adjustment (rear)
4. ON/OFF switch
5. Battery compartment
6. Telescoping rod
7. Probe storage area
8. Handle
Operating Instructions
AC CURRENT DETECTION
1. Lift the probe tip from the storage area and extend the
telescoping rod.
2. Twist the rod into the LOCKED position as indicated.
3. Slide the ON/OFF switch to the ON position. The green ON light
will illuminate and the red detection light may flash.
4. Adjust (decrease) the sensitivity adjustment until the red flashing
voltage indicator light and the beeper just stop.
5. Move the probe tip near the current carrying conductor until the
tip flashes and beeper sounds.
6. Slowly adjust the sensitivity and reduce the distance between the
probe and conductor to pinpoint and identify the conductor.
7. To turn off the audible beeper, slide the ON/OFF switch to the
position.
NOTE: RF signals in close proximity to the detector may cause the light
and beeper to latch into a constant tone and light indication. Wait until
the RF signal has disappeared before proceeding with detection.
Battery Installation
1. Unscrew the probe tip..
2. Insert two 1.5V button batteries (observe polarity).
NOTE: If your meter does not work properly, check the battery to make
sure that it still good and that it is properly inserted.
You, as the end user, are legally bound (Battery
ordinance) to return all used batteries and accumulators;
disposal in the household garbage is prohibited!
You can hand over your used batteries / accumulators,
gratuitously, at the collection points for our branches in
your community or wherever batteries / accumulators
are sold!
Disposal
Follow the valid legal stipulations in respect of the
disposal of the device at the end of its lifecycle
Copyright © 2008 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in
any form

Summary of content (1 pages)