USER MANUAL ENVIRONMENTAL METER Model EN510 Additional User Manual Translations available at www.extech.
Introduction Thank you for selecting the Extech EN510 Environmental Meter. This instrument measures Air Velocity with Air temperature, Air Flow (volume), Light, Relative Humidity % with Air temperature, Dew Point temperature, Web bulb temperature, Type K temperature (external probe), Heat Index temperature, and Wind Chill temperature. The backlit LCD includes primary and secondary displays plus a variety of intuitive status indicators.
Meter Description 1. Air Velocity Vane 2. Type K Thermocouple Input 3. LCD Display 4. Display Hold / Mode Button 5. Light Sensor 6. Humidity/Temperature sensors 7. Power/Backlight button 8. MAX-MIN Record/Enter Button Lanyard not pictured. Battery compartment and tripod mount on rear of meter LCD Description 1. MAX/MIN Record icon 2. MAX Display 3. MIN Display 4. Display Hold 5. Primary Display 6. Units of Measure 7. Temperature Display 8. Reverse Orientation Display 9.
Meter Functions Meter Power • • • Three (3) 1.5V ‘AAA’ batteries power the meter. The battery compartment is located on the rear of the meter secured by one flat head screw. Short press the power button to switch the meter ON. The meter will begin displaying readings for the mode selected. Long press the power button to power OFF the meter. Auto Power OFF (APO) In order to conserve battery life, the meter will automatically switch off after approximately 10 minutes of inactivity.
Reversible Backlight LCD The LCD display automatically reverses its presentation orientation depending on measurement mode. This is to match the display orientation with the selected sensor mode (light, air velocity, or environmental measurements such as Wet bulb temperature). The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. With the meter powered ON, short press the power button to turn ON the backlight. The backlight will power OFF after several seconds.
Air Flow (CFM, CMM) Volume Measurements • Press and hold the M button until ‘AirFl’ appears. • Enter the duct area (in square feet or meters) by short pressing M and long pressing R until ‘m-2’ (m2) or ‘f-2’ (ft2) appears. Short press the power button to increase or H to decrease the area; Long press the power or H button to scroll quickly. See area equations below.
Maintenance Battery Replacement When the low battery icon appears on the display please replace the batteries as described below. 1. Remove power from the meter. 2. Remove the flat head screw that secures the battery compartment at the back of the meter. 3. Open the battery compartment and replace the three (3) 1.5V ‘AAA’ batteries observing correct polarity. Re-assemble the meter before use.
Specifications General Display Backlit LCD 3.0 x 3.5 cm (1.18 x 1.
Relative Humidity plus Air Temperature %RH 10 to 95 0.1 ±(4%RH) @ < 70%RH ±(4%rdg +1.2%RH) @ > 70%RH °C 0 to 50 0.1 ±1.2°C °F 32 to 122 0.1 ±2.5°F Light (automatic ranging) Lux Ft-cd 0 to 2,200 1 1,800 to 20,000 10 0 to 204.0 0.1 170 to 1,860 1 0.024 to 36000 0.001 / 0.01 / 0.1 / 1 0.847 to 1271300 0.001 / 0.01 / 0.1 / 1 / 10(x10) / 100(x100) ±(5%rdg + 8dgts) Air Flow CMM (m3) 3 CFM (ft ) Dew Point Temperature Unit Range Resolution °C -25.3 to 49.0 0.1 °F -13.5 to 120.
Wind Chill °C -9.4 to 44.2 0.1° ±2.0 °C °F 15.0 to 112.0 0.1° ±3.6 °F Wind chill value is displayed only when the temperature is < 15oC (59oF) and air velocity is > 1.4 m/s Unit of Measure Conversion Table m/s ft/min knots km/h MPH 1 196.87 1.944 3.6 2.24 1 ft/min 0.00508 1 0.00987 0.01829 0.01138 1 knot 0.5144 101.27 1 1.8519 1.1523 1 km/h 0.2778 54.69 0.54 1 0.6222 1 MPH 0.4464 87.89 0.8679 1.6071 1 1 m/s Copyright © 2017 FLIR Systems, Inc.