Specifications
Function Range Resolution Accuracy
Non-contact
AC Voltage
detector
100 to
600VAC
Resolution & accuracy do not apply since the
meter does not display the voltage in this mode.
The lamp at the top of the meter’s display flashes
when voltage is sensed.
400mV 0.1mV
±(0.5% reading + 2 digits)
4V 0.001V
40V 0.01V
400V 0.1V
±(1.0% reading + 2 digits)
DC Voltage
(V DC)
600V 1V
±(1.5% reading + 2 digits)
400mV 0.1mV
±(1.0% reading + 30 digits)
4V 0.001mV
40V 0.01V
400V 0.1V
±(1.5% reading + 3 digits)
AC Voltage
(V AC)
(50 / 60Hz)
600V 1V
±(2.0% reading + 4 digits
400μA 0.1μA ±(1.0% reading + 3 digits)
4000μA 1μA
40mA 0.01mA
400mA 0.1mA
±(1.5% reading + 3 digits)
DC Current
(A DC)
10A 0.01A
±(2.5% reading + 5 digits)
400μA 0.1μA ±(1.5% reading + 5 digits)
4000μA 1μA
40mA 0.01mA
400mA 0.1mA
±(1.8% reading + 5 digits)
AC Current
(A AC)
(50 / 60Hz)
10A 0.01A