User manual
EX36x‐en‐GB_V1.61/17
25
GENERALSPECIFICATIONS
Display 6000‐countMulti‐FunctionLCD
Over‐rangeindication “OL”or“‐OL”isdisplayed
Conversionrate 3updatespersecond
MaximumVoltage 1000VACRMSor1000VDCmaximumappliedtoanyterminal
Polarityindication Automatic:positiveimplied,negativeindicated(‐)
Lowbatteryindication
isdisplayed.Meteralsodisplaysavailablebatterypowerat
startup
AutoPowerOFF After20minutes(canbedisabledbyholdingtheM(MODE)button
whileturningthemeterON)
OperatingTemperatureandHumidity
‐10~10C(14~50F)
10~30C(50~86F);80%RHmaximum
30~40C(86~104F);75%RHmaximum
40~50C(104~122F);45%RHmaximum
StorageTemperatureandHumidity
‐20~60C(‐4~140F);80%RHmaximum(withbatteryremoved)
OperatingAltitude 2000m(6562’)
Batterypower 9Vbattery(batterylife200hourstypicalwithalkalinebatterytypes)
Fuseprotection 11A/1000V(10x38mm)AC/DChighenergy(IR20kA)DMM‐B
‐11A
(EX365only)
Weight 250g(8.8oz.)Includingbattery
Dimensions(WxHxD) 74x156x44mm(2.9x6.1x1.7”)
SafetyStandards ComplieswithEN61010‐1,CATIV600V,CATIII1000V,Pollution
Degree2
EMC EN61326‐1
ShockandVibration SinusoidalvibrationMIL‐PRF‐28800F(5~55Hz,3gmax.)
DropProtection 1.2m(4’)dropontohardwoodonconcreteflooring
ForIndoorUse
Copyright©2015‐2017FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
ISO‐9001Certified
www.extech.com