User manual

PH210-en-GB_v1.3 11/14
7
BatteryReplacement
The9Vbatterythatpowerstheinstrumentrequiresreplacingwhenthebatteryiconflasheson
thedisplay.Toreplacethebatteryfollowthestepsbelow:
1. Disconnecttheelectrode/probefromtheinstrument.
2. OpentherearbatterycompartmentusingascrewdrivertoremovethetwoPhillipshead
screwsthatsecurethecompartment.
3. Carefullyremovetheoldbatteryfromthecompartmentlidchamberandgentlyunsnap
thebattery’sterminalsfromthewired(red/black)connectors.
4. Installanew9Vbatterybysnappingthebatteryterminalsontothewiredconnectors
(observingcorrectpolarity)andinsertingthebatteryintothecompartmentlidchamber.
5. Installthecompartmentlidchamberintothemeterhousingandsecurewiththetwo
screws.
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriate
collectionsites,theretailstorewhere thebatterieswerepurchased,orwherever
batteriesaresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuserisobligatedtotake
endoflifedevicestoadesignatedcollectionpointforthedisposalofelectricalandelectronic
equipment.
ElectrodeMaintenanceandTroubleshooting
Mostmeasurementdifficultiesarecausedbytheelectrode.Carefulhandlingandproperstorageof
theelectrodewillincreasemeasurementaccuracyandelectrodelife.
Electrodesshouldberinsedoftentoremovealltracesofstoragesolution,processmedium,or
previoustestsolution.
ElectrodeStorage
The electrode should be stored in a soakerbottle or wetting capfilled withpH4buffer.Never
storeanelectrode indistilledordeionized waterasthis willcause migrationof the fillsolution
fromtheelectrode.
ElectrodeCleaning
The type of electrode coating that occurs over time will determine the cleaning method. Soft
coatings can be removed by vigorous stirring or by using a squirt bottle. Organic chemical
coating(hard coatings)shouldbe chemicallyremoved.Onlyinextremecasesshouldthebulbbe
cleanedmechanically.