Warning!
For GROUP II applications ONLY, the user shall not carry more than a
maximum of 2 spare Duracell Plus MN1500 or Duracell Procell MN1500
cells into the hazardous area. Precautions shall be taken to prevent short
circuiting of un-protected cell terminals. Discharged cells shall be
removed from the hazardous area.
Warning! For Class 1 Division 1 applications ONLY, the user shall not carry more
than a maximum of 2 spare Duracell Plus MN1500, Duracell Procell
MN1500/PC1500 or Energizer E91 cells into the hazardous area.
Precautions shall be taken to prevent short circuiting of un-protected
cell terminals. Discharged cells shall be removed from the hazardous
area.
Warning!
For GROUP II applications ONLY the rated ambient temperature range of
the equipment is from -20
o
C to 60
o
C.
Warning!
For GROUP I applications ONLY the rated ambient temperature range of
the equipment is from 0
o
C to 60
o
C.
Warning!
For Class 1 Division 1 applications ONLY the rated ambient temperature
range of the equipment is from -20
o
C to 55
o
C/50
o
C dependent on the
cells used (see Table 1).
Warning! If any damage to the iCAM501U is found which may invalidate the
required safety minimum IP rating (IP54 for group I applications), (IP20
for group II and class 1 division 1 applications), the iCAM501U shall be
removed from the hazardous area.
Important Any repairs or replacement of parts MUST be performed by the
manufacturer or its nominated sub contractor or agent.
Important The iCAM501U contains no user serviceable parts. The main part of the
enclosure must not be opened under any circumstances.
Application Duracell Plus
MN1500
Duracell
Procell
MN1500
Duracell
Procell
PC1500
Energizer E91
ATEX GROUP
II
OK, T4 at 60
o
C OK, T4 at 60
o
C
CANNOT BE
USED
CANNOT BE
USED
ATEX GROUP
I
OK, T4 at 60
o
C OK, T4 at 60
o
C
CANNOT BE
USED
CANNOT BE
USED
CLASS 1
DIVISION 1
OK, T4 at 55
o
C OK, T4 at 55
o
COK, T3C at
50
o
C
OK, T3C at
50
o
C
Table1iCAM501Ucellusagetable