Safety Manual

Charging Instructions & Safety Manual
MydCombi Base Unit
Page 7 of 9
At 0°, 45°, 90°, 135°, 180°, 225°, 270°,
and 315°
At 0°, 45°, 90°, 135°, 180°, 225°, 270°,
and 315°
Voltage Dips > 95 reduction, 1 period
At
Voltage Dips > 95 reduction, 1 period
At
Voltage Interruptions > 95% reduction,
250/300 periods
Voltage Interruptions > 95% reduction,
250/300 periods
Immunity to RF Wireless Communications Equipment
Test Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Max. Power
(W)
Distanc
e (m)
Immunity test
level (V/m)
385
380 390
TETRA 400
Pulse
b)
18 Hz
1.8
0.3
27
450
430 470
GMRS 460, FRS
460
FM
c)
± 5 kHz
Deviation 1
kHz sine
2
0.3
28
710
704 787
LTE Band 13, 17
Pulse
b)
217 Hz
0.2
0.3
9
745
780
810
800 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850, LTE
Band 5
Pulse
b)
18 Hz
2
0.3
28
870
930
1720
1 700 1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4, 25;
UMTS
Pulse
b)
217 Hz
2
0.3
28
1845
1970
2450
2 400 2 570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE Band
7
Pulse
b)
217 Hz
2
0.3
28
5240
5 100 5 800
WLAN 802.11
a/n
Pulse
b)
217 Hz
0.2
0.3
9
5500
5785
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because, while it does not
represent actual modulation, it would be worst case.
Bluetooth Module
Designation
Unit
Value
Transmit and receive frequencies
MHz
2402-2480
Antenna(s) & Gain
dBi
Internal Antenna, Gain: 0.5dBi
Transmission power
mW
0.79