Owner's manual

Section 3: Programming
01/12 43 50647 Rev. 9
3.5.1. I/O MENU DESCRIPTIONS, CONTINUED
EXAMPLE
The required ID is 3 (decimal equivalent is 51).
1. Press the ID key.
Either idnonE or the current ID shows in the large display.
2. Using the numeric keypad, enter 51.
3. Press ENTER.
When the PC sends ID 3, then a upper case W followed by a C carriage
return (
3, W, Cr), the instrument with that ID responds with a data
transmission.
SEtPnt
The operation of Setpoint requires that accessories 5806 (Fiber Optic Cable)
and 2850 (Safe Area Relay Box) be installed. There are five modes of setpoint
operation that may be selected from during the installation and programming by your
local Fairbanks Service Center or Authorized Factory trained Distributor.
SEtPnt Modes
PROMPT CHOICE(S)
SPCont
Continuous operation of setpoints, ON at entered target weight.
MUST press START to begin function.
SPChEC
Check weighing, such as UNDER/ACCEPT/OVER, with setpoint
Relays 1 and 2 active for alarms.
SPALrt
Setpoint Relay 8 is active, used as both LOW and HIGH alarms in
SoLo mode.
Or Relays 4 and 8 if in Paired mode.
SPAUto
Automatic batching, a range of materials can be added one at a time to
complete a batch.
SPAUto will not function when in the Gallons mode.
When configuring SdELAY, to perform the delay in minutes and
seconds between ingredients, the setting is located in the main I/O
Menu.
SPhAnd
This operation requires a manual START at each setpoint.